Please use this identifier to cite or link to this item: doi:10.22028/D291-47470
Title: Cardiac Implantable Electronic Device Lead Perforation: A 25-Year Single-Center Experience
Author(s): Al-Maisary, Sameer
Karck, Matthias
Guzman-Ruvalcaba, Mario Jesus
Arif, Rawa
Romano, Gabriele
Language: English
Title: Journal of Clinical Medicine
Volume: 15
Issue: 7
Publisher/Platform: MDPI
Year of Publication: 2026
Free key words: cardiac implantable electronic device
lead perforation
computed tomography
DDC notations: 610 Medicine and health
Publikation type: Journal Article
Abstract: Background: Cardiac implantable electronic device (CIED) lead perforation is a rare but potentially catastrophic complication. As global device implantations increase, understanding the clinical spectrum and optimal management of this complication is essential. This study characterizes the clinical presentation, diagnostic strategies, and outcomes of lead perforation over a 25-year period. Methods: A retrospective analysis was conducted on 32 patients diagnosed with CIED lead perforation between 2000 and 2025 at a high-volume center. Perforations were classified by timing: acute (<24 h), subacute (1–30 days), and chronic (>30 days). Data included demographics, comorbidities, imaging modalities, and procedural interventions. Results: The mean patient age was 76.0 ± 11.7 years, with a mean body mass index (BMI) of 25.5 ± 3.4 kg/m2. Subacute presentation was the most frequent (59.3%, n = 19), followed by acute (28.1%, n = 9) and chronic (12.5%, n = 4) cases. The right ventricle was the primary site of perforation (90.6%). While chest X-rays served as an initial screening tool in 62.5% of cases, diagnosis relied on multimodal imaging, with Computed Tomography (CT) providing definitive confirmation in 31.3% of the cohort, particularly when lead parameters remained stable. Management was risk-stratified based on hemodynamic status. The majority of patients (71.9%, n = 23) underwent successful transvenous lead removal via simple traction. However, 25% (n = 8) presented with hemodynamic instability, and 21.9% (n = 7) suffered from cardiac tamponade. These high-risk cases required surgical intervention, including sternotomy (n = 4), thoracotomy (n = 2), or pericardiotomy (n = 3). Notably, 62.5% of hemodynamically unstable patients were on oral anticoagulants. All patients survived to discharge, with no in-hospital mortality. The median length of hospital stay was 3 days. Conclusions: CIED lead perforation often presents subacutely with subtle clinical signs. CT imaging has emerged as the gold standard for definitive diagnosis. While percutaneous transvenous removal is safe and effective for stable patients, immediate surgical backup is vital, as patients—particularly those on anticoagulation—can deteriorate rapidly.
DOI of the first publication: 10.3390/jcm15072705
URL of the first publication: https://doi.org/10.3390/jcm15072705
Link to this record: urn:nbn:de:bsz:291--ds-474707
hdl:20.500.11880/41517
http://dx.doi.org/10.22028/D291-47470
ISSN: 2077-0383
Date of registration: 15-Apr-2026
Description of the related object: Supplementary Materials
Related object: https://www.mdpi.com/article/10.3390/jcm15072705/s1
Faculty: M - Medizinische Fakultät
Department: M - Chirurgie
Professorship: M - Keiner Professur zugeordnet
Collections:SciDok - Der Wissenschaftsserver der Universität des Saarlandes

Files for this record:
File Description SizeFormat 
jcm-15-02705.pdf924,88 kBAdobe PDFView/Open


This item is licensed under a Creative Commons License Creative Commons